
SECTION 12—MOTORS
DETAIL “A” - REMOVE END CAP
End Cap Screws | End Cap |
Screwdriver to Remove
End Cap Screws (STEP 1)
DETAIL “B” - LOCATE MOTOR BRUSH
Spring Retainer
Motor Brush
Brush Housing
Shunt Wire
NOTE: Brush assembly on one side of Motor shown.
DETAIL “C” - REMOVE SHUNT WIRE
Motor Brush
Shunt Wire
Screwdriver to
Remove Shunt Wire
Shunt Wire Mounting Screw Screw (STEP 4)
DETAIL “D” - REMOVE LARGE BRUSH
Screwdriver to Release Tension | Spring Retainer | |
on Spring Retainer (STEP 5) | ||
|
Shunt
Wire
Motor
Brush
FIGURE 12.4 Replacing Motor Brushes
Electro-Mechanical Parking Brake Testing
NOTE: For this procedure, refer to FIGURE 12.5 on page 119.
NOTE: This procedure should only be performed on wheelchairs with conventional motor/gearbox assembly.
1.Engage the motor release lever. Refer to Engaging/Disengaging Motor Release Lever on page 114.
2.On the
3.Set the digital multimeter to read ohms.
4.Measure the resistance between the two brake contacts.
•A normal reading is between
•A reading of 0 ohms (W) or a very high reading; i.e., MEG ohms or O.L. (out of limit) indicates a shorted brake or an open connection respectively. If either condition exists, send the motor to Invacare Technical Service for inspection/repair.
Pronto® Series | 118 | Part No. 1125075 |