SECTION 3—SAFETY INSPECTION/TROUBLESHOOTING
Charger Digital
Port Voltmeter
BATTERY
QUALITY
0 to 2 volts | Good |
|
|
2 to 2.5 volts | Poor |
|
|
2.5 or more | Bad |
|
|
FIGURE 3.1 Field Load Test
Motor Testing
NOTE: For this procedure, refer to FIGURE 3.2.
1.On the
2.Set the digital multimeter to measure resistance (ohms).
3.Measure the resistance between the two motor contacts.
Cap
Motor
NOTE: A normal reading is between .5 and 5 ohms. A reading of 0 ohms or in excess of 15 ohms indicates a problem. High readings are generally caused by bad connections and/or damaged brushes. Contact Invacare.
Part No. 1125075 | 31 | Pronto® Series |