TDS1000/2000-Series Digital Oscilloscope User Manual 89
NOTE. The oscilloscope cancapture a waveform in normal YTmode
at any sampling rate. You can view the same waveform in XY mode.
Todo so, stop the acquisition and change the display format to XY.
The controls operate as follows:
HThe channel 1 VOLTS/DIV and VERTICAL POSITIONcontrols
set the horizontal scale and position.
HThe channel 2 VOLTS/DIV and VERTICAL POSITIONcontrols
continue to set vertical sca le and position.
The following functions do not work in XY display format:
HReference or Math waveforms
HCursors
HAutoset (resetsdisplay format to YT)
HTime base controls
HTrigger controls