Reference
TDS1000/2000-Series Digital Oscilloscope User Manual 111
Key Points
Self Calibration. The self calibration routineoptimizes the oscilloscope
accuracy for the ambient temperature. Formaximum accuracy, you
should perform a self cal if the ambient temperat ure changes by 5°C
or more. Follow the directions on the screen.
System Status
Selecting System Statusfrom the Utility Menu displays the menus
available for obtaining a list of controlsettings for each group of
oscilloscope controls.
Push any front-panel menu button to remove the status screen.
Options Comments
Horizontal Lists horizontalparameters of channels
Verti cal Lists vertical parameters of channels
Trigger Liststrigger parameters
Misc Lists the model of the oscilloscope and the software version
number
If the TDS2CMA module is installed, lists valuesof the
communicationsparameters