Application Examples
TDS1000/2000-Series Digital Oscilloscope User Manual 59
To set up to measure propagation delay, follow these steps:
1. If the channels are not displayed, push the CH 1 MENU and t hen
CH 2 MENU buttons.
2. Push the AUTOSET button to trigge r a stable display.
3. Adjust the horizontaland vertical controls to optimize the
display.
4. Push the CURSOR button to see the Cursor Menu.
5. Push the Typ e option button and select Time.
6. Push the Source option button and select CH1.
7. Turn the CURSOR 1 knob to place the c ursoron the act ive edge
of the chip-select signal.
8. Turn the CURSOR 2 knob to plac e the second cursor on the data
output transition.
9. Read the propagation delay in the Delta readout in the Cursor
Menu.