Reference
TDS1000/2000-Series Digital Oscilloscope User Manual 79
AutosetWhen you push the AUTOSET button, the oscilloscope identifies the
type of waveform and adjusts controls to produce a usable display of
the input signal.
Function Setting
Acquire mode Adjusted to Sampleor Peak Detect
Display format Set to YT
Display type Setto Dots for a video signal, set to Vectors
for an FFT spectrum; otherwise, unchanged
Horizontal position Adjusted
SEC/DIV Adjusted
Trigger coupling Adjustedto DC, Noise Reject,or HF Reject
Trigger holdoff Minimum
Trigger level Set to 50%
Trigger mode Auto
Trigger source Adjusted; refer to page 80; cannot use Autoset
on the EXT TRIG signal
Trigger slope Adjusted
Trigger type Edge or Video
Trigger VideoSync Adjusted
Trigger VideoStandard Adjusted
Verticalbandwidth Full
Verticalcoupling DC (if GND was previously selected); AC for a
video signal; otherwise,unchanged
VOLTS/DIV Adjusted