Application Examples
TDS1000/2000-Series Digital Oscilloscope User Manual 53
10. Turn the CURSOR 2 knob to plac e the second cursor at the point
where the waveform crosses the second graticule line above
center screen. This is the 90% leve l of the waveform.
11. The Delta readout in t he Cursor Menu is the rise time of the
waveform.
5 divisions
NOTE. The Rise Time measurement isav ailable as anaut omatic
measurement in the Measure Menu, described on page 94.
The Rise Time measurement also displayswhen you select the Rising
Edge option in the AUTOSET Menu. Refer to page 82.