TDS2CMA Communications Module
130 TDS1000/2000-Series Digital Oscilloscope User Manual
Checking Module Installation
To check that the module is correctly insta lled, power on the
oscilloscope. The power-on screen should list the TDS2CMA module
and include the message Power-up Tests Passe d.If the oscilloscope
does not recognize the module at power on, follow the steps in
Troubleshooting Module Installation.
Troubleshooting Module Installation
If the oscilloscopedoes not recognize the module at power on,
follow these steps:
1. Power offthe oscilloscope.
2. Follow the ESD precautions on page 128.
3. Disconnect all cables from the module.
4. Remove the module as described on page 129.
5. Examine the oscilloscopeconnec tor for bent, broken, or missing
pins. If any pins are bent, carefully strai ghten them.
6. Reinstall the module onto the oscilloscope.
7. Power on the oscilloscope. If the oscilloscope still does notshow
the module installed, contact the ne arest Tektronix service center.