TDS1000/2000-Series Digital Oscilloscope User Manual 151
Appendix A: SpecificationsAll specifications apply to the T DS1000-and T DS2000-series
oscilloscopes.P2200 probe speci fications appear at the end of this
chapter. Toverify that the oscilloscope meets specifications,the
oscilloscope mustfirst meet the following conditions:
HThe oscilloscope musthave been operating continuously for
twenty minutes withinthe specified operating temperature.
HYou must perform the Do Self Cal operation, accessible through
the Utility menu, if the operating temperaturechanges by more
than 5°C.
HThe oscilloscope mustbe within the factory calibration interval
All specifications are gua ranteed unless noted “typical.”
OscilloscopeSpecifications
Acquisition
Acquisition Modes Sample, Peak Detect, and Average
Acquisition Rate,
typical
Up to 180 waveforms per second, per channel (Sample
acquisition mode,no measurements)
Single Sequence Acquisition Mode Acquisition Stops After
g
q
Sample, Peak Detect Single acquisition, all channels
simultaneously
Average N acquisitions,all channels
simultaneously,N is selectable
from 4, 16, 64, and 128