Reference
TDS1000/2000-Series Digital Oscilloscope User Manual 103
Coupling. Coupling allows youto filter the trigger signal used to
trigger an acquisition.
Option Details
DC Passes all componentsof the signal
Noise Reject Adds hysteresis to the trigger circuitry;this reduces sensitivity
which reduces the chance of falselytriggering on noise
HF Reject Attenuatesthe high-frequency componentsabove 80 kHz
LF Reject Blocksthe D C componentand attenuates the low -frequency
components below 300 kHz
AC Blocks DC components and attenuates signals below 10 Hz
NOTE. Trigger coupling affects only the signal passed to the trigger
system. It does not affect the bandwidth or coupling of the signal
displayed on the screen.
Pretrigger. The trigger position is typically setat the horizontal center
of the screen. In this case, you are able to view five divisions of
pretrigger information.Adjusting the Horizontal Position of the
waveform allows you to see more or less pretrigger informat ion.