Application Examples
TDS1000/2000-Series Digital Oscilloscope User Manual 61
6. Push the Source option button to select CH1.
7. Turn the TRIGGER LEVEL knob to set the trigger l evel near
the bottom of the signal.
8. Push the When option button to select =(equal).
9. Push the Set Pulse Width option button, and t urn the USER
SELECT knob to set the pulse width to the value reported by t he
Pulse Widthmeasurement in step 3.
10. Push -- m o r e -- p a g e 1 o f 2 and set the Mode option to Normal.
You should achieve a stable display with the oscilloscope
triggering on normal pulses.
11. Push the When option button to select ΒΈ,<,or>. If there are
any aberrant pulses that meet t he specified When condition, the
oscilloscope triggers.
NOTE. The trigger frequency readout shows the frequency of events
the oscilloscope considers to be triggers, and may be lessthan the
frequency of the input signal in Pulse Width trigger mode.