Calibration

You can calibrate the Ethernet/RS-232 Interface for:

Voltage program

Voltage readback

Current program

Current readback

Overvoltage protection

The following equipment will be required to accurately calibrate your unit:

Digital Voltmeter, 5 1/2 digit, 0.1% accuracy or better, with test leads

Current sensing shunt resistor, rated for 150% of maximum output current, 0.25% accuracy or better

Connection wires rated for the unit's maximum output current and voltage

A network ready, or RS-232 equipped computer to send the calibration commands to the unit

Calibrate the unit according to the following procedures, referring to “Command Reference” for more information about the calibration commands used.

The calibration procedures in this section are designed to be performed at an ambient temperature of 25°C ± 5°C.