Current Mode Calibration

Current Mode Calibration

Current Mode Calibration

Current Calibration Setup

1.Disconnect the load from the power supply to be calibrated.

2.Connect a shunt across the supply's output terminals.

3.Connect a voltmeter across the shunt.

Power Supply

Power Supply

Positive Output

Negative Output

Current Sensing

Resistor

V

+ -

VOLTMETER

Figure 4-2Current Calibration Setup

Current Program Calibration Procedure

1.Connect the shunt and voltmeter to the power supply as shown in Figure 4-2.

2.Activate calibration mode by sending command CMODE ON or CMODE 1 to the power supply.

3.Send command ILO; VLO to the power supply. Measure and record the output shown on the external voltmeter.

4.Send command IHI; VHI to the supply Measure and record the output voltage shown on the external voltmeter.

5.Calculate ILO and IHI from the voltages read from the external voltmeter and the shunt resistance. I=V/R.

6.Send the command IDATA <ilo>,<ihi> to the power supply. <ilo> and <ihi> are the current values obtained from sending the ILO and IHI commands to the power supply. When the power supply is calibrated, the low to high current program calibration values are stored as constants.