Calibration

Voltage Readback Calibration Procedure

1.Set the power supply for calibration as in Figure 4-1.

2.Activate calibration mode by sending command CMODE ON or CMODE 1 to the power supply.

3.Send command VRLO; IRLO to the power supply. Wait for the supply to settle. Measure and record the output shown on the external voltmeter. Send VRLO again.

4.Send VRHI; IRHI to the supply. Wait for the supply to settle. Measure and record the output voltage shown on the external voltmeter. Send VRHI again.

5.Send the command VRDAT <vlo>,<vhi> where <vlo> and <vhi> are the values read from the voltmeter after the VRLO and VRHI commands were sent. The processor calculates the offset value required to calibrate the power supply. When the power supply is calibrated, the low to high voltage readback calibration values (offsets) are stored as constants.

6.Use commands VSET and VOUT? commands to confirm that the calibration was successful and that linearity is observed. Refer to the voltage readback accuracy specification in Appendix A.

7.Turn off calibration mode by sending the command CMODE OFF or CMODE 0 to the power supply.

4–4

TM-XR9B-01XN

Page 78
Image 78
Xantrex Technology ENET-XFR3 manual Voltage Readback Calibration Procedure