Table 3-7Command Reference

 

 

Command

Description

IMAX <current>

S ts an upper soft limit on the supply’s programmed output current. If the soft limit

 

is

xceeded, or if the soft limit value is lower than the present output current

 

setting, the supply will ignore the command, turn on the ERR LED, and set the

 

ERR b in the bit registers.

 

Range: 0 to model maximum output current (IMAX)

 

Initial value: model IMAX

IMAX?

Asks for he supply’s soft current limit setting.

 

R

ponse: IMAX <current>

IOUT?

M asures the supply’s actual current output using the built-in current readback

 

circuitry.

 

R

ponse: IOUT <current>

 

 

IRDAT

Calculates a d records the slope and offset for readback voltage using IRLO and

<Ilo>,<Ihi>

IRHI data. Set CMODE ON before using this command. See also the calibration

 

procedures in Chapter 4.

 

<Ilo> and <Ihi> are in <current> format.

 

 

IRHI

The power supply outputs a current value to an external device connected as part of

 

the cal bration procedure and records a current readback value internally. These

 

values are at the high end of the programmed current range. Refer to the output

 

value as IRHI and record it to use as input with the IRDAT command. Set

 

CMODE ON before using this command. See also the calibration procedures in

 

Chapter 4.

 

 

IRLO

The power supply outputs a current value to an external device connected as part of

 

the calibration procedure and records a current readback value internally. These

 

values are at the low end of the programmed current range. Refer to the output

 

value as IRLO and record it to use as input with the IRDAT command. Set

 

CMODE ON before using this command. See also the calibration procedures in

 

Chapter 4.

 

 

ISET <current>

Sets the power supply’s output current in amps (default) or in milliamps. This

 

programmed current is the actual output in CC mode or the current limit in CV

 

mode.

 

Range: 0 to model maximum output current (IMAX)

 

Initial value: 0 amps

 

 

ISET?

Asks for the supply’s present output current setting. Does not apply to current

 

settings which are being held. See HOLD command.

 

Response: ISET <current>

 

 

 

3–12

TM-XR9B-01XN