3.4Quick Media Calibration

3.4Quick Media Calibration

+When ‘Uncalibrated’ is displayed, follow the Media Sensor Calibration procedure in Section 5.0.

At the factory, the printer is calibrated to sense a wide range of media types. Quick Media Calibration fine-tunes the media sensor for your gap, notch or reflective media application (this is not required for continuous media). Perform this calibration during initial set-up or after changing your media type. To calibrate:

ŒEnsure that media is loaded (see Section 3.1), that the Media Sensor is adjusted (see Section 3.2), and that the printer is idle.

Press and hold the FEED key. The printer will begin advancing media; allow at least one label gap or mark to advance under the sensor during this process.

Upon successful completion, the ‘Calibration Completed’ message will flash; the printer will feed to the next label TOF and ‘Ready’ will be displayed. (A ‘Warning Low Backing’ message may appear if using notched media or media on a transparent liner; however, the calibration was successful).

þNote: Media containing large gaps may require a change in the ‘Paper Out Distance’ setting; see Section 4.1.2.

Calibration Hints:

In certain cases, the printer may have trouble differentiating between the label and liner. If the printer stops feeding in the middle of a label or if ‘Cannot Calibrate’ is displayed, try calibrating over a longer distance:

Press and hold the FEED key to allow two gaps or marks to advance under the sensor.

If the printer continues to stop in the middle of a label, or if ‘Cannot Calibrate’ is displayed again:

Press and hold the FEED key to allow three or more gaps or marks to advance under the sensor.

If this method also fails, see Media Sensor Calibration, Section 5.0.