Instruction Manual

IM-106-340, Rev. 4.0

May 2006

Figure 8-9. Fault 7, High Case Temp

 

HEATER T/C

 

DIAGNOSTIC

HEATER

SW2

ALARMS

O2 CELL

ON

 

 

CALIBRATION

 

CALIBRATION RECOMMENDED

Oxymitter 4000

Fault 7, High Case Temp

Figure 8-9shows the electronic assembly for an Oxymitter 4000 with a membrane keypad (upper view) and an Oxymitter 4000 with an LOI (lower view).

Membrane Keypad

When Fault 7 is detected, The HEATER LED flashes three times, pauses for three seconds,

TEST

POINTS

INC INC

HIGH LOW GAS GAS

DEC DEC

O2 CELL mV +

O2 CELL mv - HEATER T/C + HEATER T/C -

CAL

TEST GAS + PROCESS - % O2

TP1 TP2 TP3 TP4

TP5

TP6

J1 RED YEL GRN ORG

and repeats.

1.

If the case temperature exceeds [185°F

 

(85°C)], the temperature control will shut

 

off and the 4-20 mA signal output will go to

 

the default value.

2.

This signifies that the environment where

 

the Oxymitter 4000 is installed exceeds the

 

ambient temperature requirements or that

 

heat due to convection is causing case

 

temperature to rise above the limit.

3.

Placing a spool piece between the stack

 

flange and the Oxymitter 4000 flange may

KEYPAD

Alarms

Board Temp Hi

LOI

37260025

 

eliminate this problem.

4. If a spool piece does not solve the problem,

relocation is the only solution.

LOI

When Fault 7 is detected, the LOI displays the "Board Temp Hi" message. Refer to the comments in paragraphs 1 through 4 above.

8-13

Page 105
Image 105
Emerson Process Management IM-106-340 instruction manual Alarms Board Temp Hi, Fault 7, High Case Temp