Oxymitter 4000

Instruction Manual

IM-106-340, Rev. 4.0 May 2006

Figure 4-1. Electronics Housing

Terminals with LOI

Oxymitter 4000

Electronics

Housing

Rosemount Analytical Inc.

SMART FAMILY

Orrville, OH 44667-0901

800-433-6076

HART

OXYMITTER 4000

SERIAL NO.

TAG NO.

VOLTS: 85-264 VAC

48-62 Hz WATTS:

500 VA

OUTPUT: 4-20 mA

LINE FUSE:

5 Amps

+ 4-20 -

-

+

AC

 

 

AC

 

 

N

L1

4-20 mA

 

Signal

Terminal

 

Block

Logic I/O

 

LOI

Ground Lugs

37260011

Oxymitter 4000 Configuration

Located on the microprocessor board are two switches that configure outputs (Figure 4-2). To access these switches, the LOI module must be removed. SW1 determines if the 4-20 mA signal is internally or externally powered. SW2 determines:

1.Range control, HART or LOCAL.

2.Oxygen range, 0 to 10% O2 or 0 to 25% O2. (0 to 40% O2 is also configurable only through HART/AMS.)

3.The 4-20 mA signal, at fault or power up, 3.5 mA or 21.6 mA.

Remove power before changing defaults. If defaults are changed under power, damage to the electronics package may occur.

SW1 Setting

The two settings are internally or externally powering the 4-20 mA signal. The factory setting is for the 4-20 mA signal to be internally powered.

4-2

Page 64
Image 64
Emerson Process Management IM-106-340 instruction manual Electronics Housing Terminals with LOI