Figure 6-4. Menu Tree for Local Operator Interface (LOI) on the Oxymitter 4000 (Sheet 2 of 2)

(CONTINUED FROM

SHEET 1)

SYSTEM

Calib Setup

Input/Output

Parameters

Software

Status

Analog

Digital

O2 Gas 1

_____%

O2 Gas 2

_____%

O2-Reset Vals

Yes/No

O2 Out Tracks

Yes/No

O2 Cal Intervl

____H

O2-Next Cal

____H

Gas Time

___Sec.

Purge Time

___Sec.

Auto Calib?

Yes/No

O2 Type

_______

O2 Range

______%

O2 Alarm Level

_____mA

Do O2 Trim

 

Logic IO Mode See Table 4-1

Low O2 Alarm See “Abort

Calibration” in Section 9.

Input State

Force Output

O2 Slope

____ mV/D

O2 Constant

____ mV

O2 T90 Time

0:00

Auto Tune?

Yes/No

Lockout Time

0:00

Revert Time

0:00

Luminance

______

Version

xxx

Checksum

xxx

Build Number

xxx

Build Date

xxxxxx

Test Code

xx

SW Err File

xx

SW Err Line

xx

Alarms

__________

PID Parameters

115/220

Reset Device?

Yes/No

(Cal. required after reset)

NOTE

In column four of this menu, the selections in Italics are user configurable. Bold text selections are procedures; related instructions are displayed on the LOI. All other parameters are display only.

37260018

OXYMITTER 4000 SETUP AT THE LOI

In setting up the Oxymitter 4000 from the LOI, it is best to start at the SYSTEM/Calibration Setup menu, Figure 6-4.

SYSTEM/Calibration Setup

O2 Gas #1 - Enter the high or low cal gas value (the order is not important). O2 Gas #2 - Enter the second cal gas value.

NOTE

Refer to Section 9: Maintenance and Service, for calibration instructions.

6-6

Page 78
Image 78
Emerson Process Management IM-106-340 instruction manual Oxymitter 4000 Setup AT the LOI, Do O2 Trim