Oxymitter 4000

Instruction Manual

IM-106-340, Rev. 4.0 May 2006

SPECIFICATIONS

Oxymitter Specifications

O2 Range

 

 

 

Standard

0 to 10% O2, 0 to 25% O2, 0 to 40% O2 (via HART)

Accuracy

±0.75% of reading or 0.05% O2, whichever is greater

System Response to Calibration

Initial – less than 3 seconds, T90 – less than 8

Gas

seconds

 

 

Temperature Limits

 

 

 

 

 

Process

32° to 1300°F (0° to 704°C) up to 2400°F (1300°C)

 

with optional accessories

 

Electronics Housing

-40° to 158°F (-40° to 70°C) ambient

Electronics Package

-40° to 185°F (-40° to 85°C) [Operating temperature

 

of electronics inside of instrument housing, as

 

measured by a HART communicator, Rosemount

 

Analytical Asset Management Solutions software.]

Local Operator Interface

-40° to 158°F (-40° to 70°C), [above 158°F (70°C) the

 

infrared keypad will cease to function, but the

 

Oxymitter 4000 will continue to operate properly.]

Probe Lengths

18 in. (457 mm)

 

12 ft (3,66 m)

 

3 ft (0,91 m)

 

15 ft (4,57 m)

 

6 ft (1,83 m)

 

18 ft (5,49 m)

 

9 ft (2,74 m)

 

 

Mounting and Mounting Position

Vertical or horizontal;

 

 

a spool piece, (P/N 3D39761G02), is available to

 

offset transmitter housing from hot ductwork.

 

 

 

 

Materials

 

 

 

 

 

Probe

Wetted or welded parts - 316L stainless steel (SS)

 

Non-wetted parts - 304 SS, low-copper aluminum

 

 

 

Electronics Enclosure

Low-copper aluminum

 

Calibration

Manual, semi-automatic, or automatic

Calibration Gas Mixtures

0.4% O2, Balance N2

 

Recommended

8% O2, Balance N2

 

Calibration Gas Flow

2.5 l/m (5 scfh)

 

 

Reference Air

2 scfh (1 l/m), clean, dry, instrument-quality air

 

(20.95% O2), regulated to 34 kPa (5 psi)

Electronics

NEMA 4X, IP66 with fitting and pipe on reference

 

exhaust port to clear dry atmosphere

 

 

 

Electric Noise

EN 61326-1, Class A

 

Certifications

General Purpose

 

 

 

C

US

APPROVED

 

 

Line Voltage

90-250 VAC, 48/62 Hz. No configuration necessary.

 

3/4 in. -14 NPT conduit port

Table continued on next page

1-16

Page 34
Image 34
Emerson Process Management IM-106-340 instruction manual Oxymitter Specifications