Instruction Manual

Oxymitter 4000

 

 

 

IM-106-340, Rev. 4.0

 

 

 

 

May 2006

Figure 8-1. O2

Sensor mV

 

 

 

 

 

Reading vs. % O2 at 1357°F

 

 

 

 

 

(736°C) (Reference Air, 20.9%

 

 

 

 

 

O2)

 

 

 

 

 

 

 

 

200

 

 

 

 

 

 

 

 

 

O2 Sensor Performance at 1357OF

(736OC)

 

 

150

 

 

 

 

 

 

(mV)

100

 

 

 

 

 

 

EMF

 

 

 

 

 

 

 

 

 

 

 

 

 

 

50

 

 

 

 

 

 

 

0 0

0.01

0.1

1

10

100

37260043

 

 

 

Concentration O2 (%)

 

 

 

 

 

 

 

 

O2 %

100

20

15

10

9

8

7

6

5

4

EMF(mV)

-34

1.0

7.25

16.1

18.4

21.1

23.8

27.2

31.2

36.0

 

 

 

 

 

 

 

 

 

 

 

O2 %

3

2

1

0.8

0.6

0.5

0.4

0.2

0.1

0.01

EMF(mV)

42.3

51.1

66.1

71.0

77.5

81.5

86.3

101.4

116.6

166.8