Oxymitter 4000

Figure 8-14. Fault 12, EEprom Corrupt

 

HEATER T/C

 

DIAGNOSTIC

HEATER

SW2

ALARMS

O2 CELL

ON

 

 

CALIBRATION

 

CALIBRATION RECOMMENDED

Instruction Manual

IM-106-340, Rev. 4.0

May 2006

Fault 12, EEprom Corrupt

Figure 8-14shows the electronic assembly for an Oxymitter 4000 with a membrane keypad (upper view) and an Oxymitter 4000 with an LOI (lower view).

Membrane Keypad

When Fault 12 is detected, the O2 CELL LED flashes four times, pauses for three seconds, and repeats.

TEST

POINTS

INC INC

HIGH LOW GAS GAS

DEC DEC

O2 CELL mV +

O2 CELL mv - HEATER T/C + HEATER T/C -

CAL

TEST GAS +

PROCESS -

% O2

TP1 TP2 TP3 TP4

TP5

TP6

J1 RED YEL GRN ORG

1.

This alarm can occur if the EEprom is

 

changed for a later version. At power up,

 

the EEprom is not updated.

2.

To correct this problem, power down and

 

then restore power. The alarm should clear.

3.

If the alarm occurs while the unit is running,

 

there is a hardware problem on the

 

microprocessor board.

4.

If cycling the power does not clear the

 

alarm, see "Electronic Assembly

 

Replacement" in Section 9: Maintenance

KEYPAD

Alarms

EEprom Corrupt

LOI

37260030

 

and Service.

LOI

When Fault 12 is detected, the LOI displays the "EEprom Corrupt" message. Refer to the comments and procedures in paragraphs 1 through 4 above.

8-18

Page 110
Image 110
Emerson Process Management IM-106-340 instruction manual Alarms EEprom Corrupt, Fault 12, EEprom Corrupt