Figure 9-2. Inside Right Cover

ALARMS

LED

FLASHES

STATUS

 

1

OPEN

 

2

SHORTED

HEATER T/C

3

REVERSED

 

 

4

A/D COMM

 

ERROR

 

 

 

1

OPEN

 

2

HIGH HIGH

 

TEMP

 

 

HEATER

3

HIGH CASE

TEMP

 

 

4

LOW TEMP

 

5

HIGH TEMP

 

1

OPEN

O2 CELL

3

BAD

 

 

 

4

EPROM

 

CORRUPT

 

 

 

1

INVALID SLOPE

CALIBRATION

2

INVALID

 

CONSTANT

 

 

 

3

LAST CAL

 

FAILED

 

 

MANUAL

CALIBRATION

*PLACE CONTROL LOOP IN MANUAL

*IF CAL LED ON GO TO STEP 2

1

PUSH CAL

CAL LED ON

 

2

PUSH CAL

CAL LED FLASH

 

3APPLY TG1 PUSH CAL

4CAL LED ON SOLID WAIT FOR FLASH

5REMOVE TG1 & APPLY TG2

PUSH CAL

CAL LED ON SOLID

6WAIT FOR FLASH

2 FLASH-VALID CAL

3 FLASH-INVALID CAL

7REMOVE TG2

PUSH CAL

CAL LED ON FOR

8PURGE TIME CAL LED OFF

SW2 DIP SWITCH

HART 0-10% 4mA

NOT USED

LOCAL 0-25%

20mA

NOT USED

29770005

2.Verify the calibration gas parameters are correct per "Calibration with Keypad".

3.If performing a manual calibration with the CALIBRATION RECOMMENDED LED off and the CAL LED off, start at step a.

4.If performing a manual calibration with the CALIBRATION RECOMMENDED LED on and the CAL LED on, start at step b.

a.Push the CAL key. The CALIBRATION RECOMMENDED LED will come on and the CAL LED will be on solid. If a multimeter is attached across TP5 and TP6, the reading will display the percentage of oxygen seen by the cell.

b.Push the CAL key. The CALIBRATION RECOMMENDED LED will turn off and the CAL LED will flash continuously. The Oxymitter 4000 can be configured so that the 4-20 mA signal will hold the last value. The default condition is for the output to track. A flashing LED indicates that the Oxymitter 4000 is ready to accept the first calibration gas.

c.Apply the first calibration gas. (Electronics will abort the calibration if step 4 is not done within 30 minutes).

d.Push the CAL key; the CAL LED will be on solid. A timer is activated to allow the calibration gas adequate time to flow (default time of five minutes). When the timer times out, the Oxymitter 4000 has taken the readings using the first calibration gas and the CAL LED will flash continuously. The flashing indicates the Oxymitter 4000 is ready to take readings using the second calibration gas.

9-4

Page 120
Image 120
Emerson Process Management IM-106-340 instruction manual Alarms, Manual Calibration