Oxymitter 4000

Instruction Manual

IM-106-340, Rev. 4.0 May 2006

QUICK START GUIDE FOR OXYMITTER 4000 SYSTEMS

Before using the Quick Start Guide, please read "WHAT YOU NEED TO KNOW BEFORE INSTALLING AND WIRING A ROSEMOUNT ANALYTICAL OXYMITTER 4000 OXYGEN TRANSMITTER" on the preceding page.

1.Install the Oxymitter 4000 in an appropriate location on the stack or duct. Refer to "Selecting Location" in Section 2: Installation, for information on selecting a location for the Oxymitter 4000.

2.If using an SPS 4001B, connect the calibration gasses to the appropriate fittings on the SPS 4001B manifold.

3.Connect reference air to the Oxymitter 4000 or SPS 4001B, as applicable.

4.If using an SPS 4001B, make the wiring connections as shown in the SPS 4001B Single Probe Autocalibration Sequencer Instruction Manual.

5.If NOT using an SPS 4001B, make the following wire connections as shown in Figure 3: line voltage, 4-20 mA, and logic I/O.

6.Verify the Oxymitter 4000 switch configuration is as desired. Refer to "Oxymitter 4000 Configuration", "SW1 Setting", and "SW2 Setting" all in Section 3: Configuration of Oxymitter 4000 with Membrane Keypad, or "Oxymitter 4000 Configuration", "SW1 Setting", and "SW2 Setting" all in Section 4: Configuration of Oxymitter 4000 with LOI.

7.Apply power to the Oxymitter 4000; the cell heater will turn on. Allow approximately one half hour for the cell to heat to operating temperature. Once the ramp cycle has completed and the Oxymitter 4000 is at normal operation, proceed with step 8 or 9.

8.If using an SPS 4001B, initiate a semi-automatic calibration.

9.If NOT using an SPS 4001B, perform a manual calibration. Refer to "Calibration with Keypad" or "Calibration with LOI" both in Section 9: Maintenance and Service, in this instruction manual.

NOTE

If your system has a membrane keypad you can refer to the Quick Start Guide on the following pages.

viii

Page 16
Image 16
Emerson Process Management IM-106-340 instruction manual Quick Start Guide for Oxymitter 4000 Systems