Instruction Manual

IM-106-340, Rev. 4.0 May 2006

Oxymitter 4000

Figure 9-10. Cell Replacement Kitl

ANSI

Gasket

 

 

Wire and

 

 

Pad Assembly

Probe Tube

 

Anti-Seize

 

Compound

(Not Included in Kit)

 

 

 

 

22 Ga.

Corrugated

 

Wire

 

 

Seal

Cell and

 

 

 

 

Flange

Closed end

 

Assembly

 

Connector

 

 

 

Socket Head

Set Screws

 

 

 

Cap Screws

Teflon

 

 

Calibration Gas

 

Tubing

Passage

 

 

35830009

Use heat-resistant gloves and clothing when removing the probe. Do not attempt to work on these components until they have cooled to room temperature. Probe components can be as hot as 572°F (300°C). This can cause severe burns.

Disconnect and lock out power before working on any electrical components. There is voltage of up to 115 VAC.

Do not remove the cell unless certain it needs to be replaced. Removal may damage the cell and platinum pad. Go through the complete troubleshooting procedure to make sure the cell needs to be replaced before removing it.

1.Follow the instructions in "Removal and Replacement of Probe" to remove the Oxymitter 4000 from the stack or duct.

NOTE

To determine if the diffuser needs to be replaced, refer to "Calibration with

Keypad".

2.If the probe uses the standard diffusion element, use a spanner wrench to remove the diffuser assembly.

9-17

Page 133
Image 133
Emerson Process Management IM-106-340 instruction manual Cell Replacement Kitl