Instruction Manual

IM-106-340, Rev. 4.0 May 2006

Oxymitter 4000

Figure 5-2. Calibration Keys

Diagnostic

LEDs

Membrane

Keys

HEATER T/C

DIAGNOSTIC HEATER

ALARMS 02 CELL

CALIBRATION

CALIBRATION RECOMMENDED

 

02 CELL mV +

TEST

02 CELL mv -

POINTS

HEATER T/C +

 

HEATER T/C -

 

 

 

 

 

 

 

 

 

 

INC

 

INC

 

 

 

CAL

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

HIGH

 

LOW

 

 

 

 

 

 

GAS

 

GAS

 

 

 

 

 

 

 

 

 

TEST GAS +

 

DEC

 

DEC

 

 

 

 

 

 

 

 

 

 

 

 

 

PROCESS -

 

 

 

 

 

 

 

 

 

 

% 02

 

Membrane

Key

Membrane

Keys

22220023

Keys

INC and DEC. The INC and DEC keys are used to set the values of the calibration gases. Attach a multimeter across TP5 and TP6. The calibration and process gases can now be monitored. Pressing the INC or DEC once will cause the output to switch from the process gas to the calibration gas. Pressing INC or DEC a second time will increase or decrease the calibration gas parameter. If the keys have been inactive for one minute, the output reverts to the process gas. When a calibration has been initiated, the value at TP5 and TP6 is the % O2 seen by the cell.

Oxygen levels, as seen on the multimeter, are:

8.0% O2 = 8.0 volts DC

0.4% O2 = 0.4 volts DC

CAL

The CAL key can:

Initiate a calibration.

Sequence through calibration.

Abort the calibration.

NOTE

Refer Section 9: Maintenance and Service, for calibration instructions.

Model 751 Remote Powered Loop LCD Display (Optional)

Refer to Remote Powered Loop LCD manual for calibration and operation.

5-3

Page 71
Image 71
Emerson Process Management IM-106-340 instruction manual Keys, Cal, Model 751 Remote Powered Loop LCD Display Optional