Oxymitter 4000

Figure 4-2. Defaults - Oxymitter 4000 with LOI

Instruction Manual

IM-106-340, Rev. 4.0 May 2006

4-20 mA

OFF

 

ON

 

is internally

powered (Default)

4-20 mA requires

external power

HART: O2 Range set by HART/AMS (From 0 to 40% O2)

Local: O2 Range set by Pos 2

0 to 10% O2/

0 to 25% O2: O2 Range

3.5mA/21.6 mA When alarm exists, or on power up, current output goes to this value

HART 0 to 10% O2

3.5 mA

220 V

Default position (Ex-factory)

Local

0 to 25% O2

21.6mA

115 V

SW2

1

2

3

4

TP1

J1

 

TP2

 

TP3

 

TP4

RED YEL GRN ORG

 

TP5

TP6

Note:

The 115 V option at switch SW2 position 4 is active only when the heater voltage option is set to manual in the software.

37260012

LOGIC I/O

This two-terminal logic contact can be configured either as a solid-state

 

relay-activated alarm or as a bi-directional calibration handshake signal to an

 

IMPS 4000 or SPS 4001B. The configuration of this signal depends on the

 

setting of the LOGIC I/O PIN MODE via HART/AMS or LOI. The ten different

 

modes available are explained in Table 4-1.

 

Alarm

 

When configured as an alarm, this signal alerts you to an out-of-spec

 

condition. The output is +5 Vdc in series with a 340 ohm resistor.

 

For optimum performance, Rosemount Analytical recommends connecting

 

the output to a Potter & Brumfield 3.2 mA DC relay (P/N R10S-E1Y1-J1.0K).