Instruction Manual

IM-106-340, Rev. 4.0

May 2006

Figure 8-11. Fault 9, High Heater Temp

 

HEATER T/C

 

DIAGNOSTIC

HEATER

SW2

ALARMS

O2 CELL

ON

 

 

CALIBRATION

 

CALIBRATION RECOMMENDED

Oxymitter 4000

Fault 9, High Heater Temp

Figure 8-11shows the electronic assembly for an Oxymitter 4000 with a membrane keypad (upper view) and an Oxymitter 4000 with an LOI (lower view).

Membrane Keypad

When Fault 9 is detected, the HEATER LED flashes five times, pauses for three seconds, and repeats.

TEST

POINTS

INC INC

HIGH LOW GAS GAS

DEC DEC

O2 CELL mV +

O2 CELL mv - HEATER T/C + HEATER T/C -

CAL

TEST GAS + PROCESS - % O2

TP1 TP2 TP3 TP4

TP5

TP6

J1 RED YEL GRN ORG

1.

If the thermocouple produces a voltage in

 

excess of approximately 30.7 mV, the high

 

heater temp alarm activates.

2.

The 4-20 mA signal returns to the default

 

value (4 or 20 mA).

3.

This alarm is self-clearing. When

 

temperature control is restored and the

 

thermocouple voltage returns to the normal

 

range, the alarm clears.

4.

If the temperature continues to rise, the

 

next alarm will be the High High Heater

KEYPAD

Alarms

O2 Temp Hi

LOI

37260027

 

Temp alarm.

LOI

When Fault 9 is detected, the LOI displays the "O2 Temp Hi" message. Refer to the comments and procedures in paragraphs 1 through 4 above.

8-15

Page 107
Image 107
Emerson Process Management IM-106-340 instruction manual Alarms O2 Temp Hi, Fault 9, High Heater Temp