Oxymitter 4000

Figure 8-16. Fault 14, Invalid Constant

Instruction Manual

IM-106-340, Rev. 4.0

May 2006

Fault 14, Invalid Constant

Figure 8-16shows the electronic assembly for an Oxymitter 4000 with a membrane keypad (upper view) and an Oxymitter 4000 with an LOI (lower view).

HEATER T/C

DIAGNOSTIC HEATER

ALARMS O2 CELL CALIBRATION

CALIBRATION RECOMMENDED

SW2

ON

Membrane Keypad

When Fault 14 is detected, the CALIBRATION LED flashes twice, pauses for three seconds,

TEST

POINTS

INC INC

HIGH LOW GAS GAS

DEC DEC

O2 CELL mV +

O2 CELL mv - HEATER T/C + HEATER T/C -

CAL

TEST GAS + PROCESS - % O2

TP1

J1

 

TP2

 

TP3

 

TP4

RED YEL GRN ORG

 

TP5

TP6

and repeats.

1.

After a calibration has been performed, the

 

electronics calculates a cell constant value.

2.

If the cell constant value is outside of the

 

range, -4 mV to 10 mV, the alarm will

 

activate. See "Calibration with Keypad" in

 

Section 9: Maintenance and Service, and

 

verify the last calibration was performed

 

correctly.

3.

Power down the Oxymitter 4000 and

 

remove it from the stack.

KEYPAD

Alarms

O2 Cell Bad

LOI

37260032

4. Replace the cell per "Cell Replacement" in

Section 9: Maintenance and Service.

LOI

When Fault 14 is detected, the LOI displays the "O2 Cell Bad" message. Refer to the comments and procedures in paragraphs 1 through 4 above.

8-20

Page 112
Image 112
Emerson Process Management instruction manual Fault 14, Invalid Constant IM-106-340, Rev May