Oxymitter 4000

Instruction Manual

IM-106-340, Rev. 4.0 May 2006

HART COMMUNICATOR O2 CAL METHOD

Use the following procedure to perform a calibration using the HART Commu- nicator. If necessary, use the menu tree in Figure 7-3(sheet 1 of 3) for reference.

NOTE

To select a menu item, either use the up and down arrow keys to scroll to the menu item and press the right arrow key or use the number keypad to select the menu item number. To return to a preceding menu, press the left arrow key.

1.From the PERFORM O2 CAL screen, select menu item 1, O2 CAL, to access the O2 calibration procedure.

Failure to remove the Oxymitter 4000 from automatic control loops prior to performing this procedure may result in a dangerous operating condition.

2.In the first O2 CAL screen, a "Loop should be removed from automatic control" warning appears. Remove the Oxymitter 4000 from any auto- matic control loops to avoid a potentially dangerous operating condition and press OK.

3.The next several screens indicate the calibration status. At each of the following status prompts, select menu item 2, NEXT CAL STEP:

COMPLETE

CAL RECOMMENDED APPLY GAS 1

GAS 1 FLOW

4.At this point, select menu item 4, EXIT, to leave the O2 CAL procedure.

5.From the PERFORM O2 CAL screen, view menu item 3, CALSTATE, to monitor the calibration status as it updates. Or, access the O2 CALI- BRATE screen and select menu item 2, O2 CAL STATUS, to view menu item 1, CALSTATE; menu item 2, TIMEREMAIN; and menu item 3, PRESENT O2, as the calibration status updates.

6.When CALSTATE displays APPLY GAS 2, return to the O2 CAL procedure.

7.When the "Loop should be removed from automatic control" warning appears, return the Oxymitter 4000 to the automatic control loops previ- ously removed and press OK.

8.At the STOP GAS status prompt, select menu item 2, NEXT CAL STEP. When the status displays PURGING, select menu item 4, EXIT, to leave the O2 CAL procedure.

9.From the PERFORM O2 CAL screen, view menu item 3, CALSTATE, to monitor the calibration status as it updates. Or, access the O2 CALI- BRATE screen and select menu item 2, O2 CAL STATUS, to view menu item 1, CALSTATE; menu item 2, TIMEREMAIN; and menu item 3, PRESENT O2, as the calibration status updates.

10.When CALSTATE displays STOP GAS, return to the O2 CAL proce- dure.

7-8

Page 90
Image 90
Emerson Process Management IM-106-340 Hart Communicator O2 CAL Method, Complete CAL Recommended Apply GAS GAS 1 Flow