Instruction Manual

IM-106-340, Rev. 4.0 May 2006

Figure 2-5. Oxymitter 4000 Adapter Plate Installation

Oxymitter 4000

INSTALLATION FOR METAL WALL STACK OR DUCT CONSTRUCTION

INSTALLATION FOR MASONRY WALL STACK CONSTRUCTION

Mounting holes shown rotated 45o out of true position

0.50 (13)

3.75(95)

Minimum Dia hole in wall

Stack or duct metal wall

Bolt mounting plate to outside wall surface

Field weld pipe to adapter plate

Mounting Holes shown rotated 45o out of true position

Joint must be air tight

0.50 (13)

4.50(114) O.D. Ref

Pipe 4.00 in. Sched 40 Pipe Sleeve (Not by Rosemount Analytical) length by Customer

Masonry

Weld or bolt mounting plate to metal wall of stack or duct; Joint must be air tight.

Outside wall

surface

stack wall

 

Bolt adapter plate to

 

outside wall surface

2.50 (63,5)

 

Minimum Dia

 

hole in wall

 

Stack or duct

 

metal wall

 

 

Joint must

 

be air tight

Weld or bolt adapter plate to

Outside wall

surface

metal wall of stack or duct.

 

Joint must be air tight.

 

Notes: 1. Dimensions are in inches with millimeters in parentheses.

2.All masonry stack work and joints except adaptor plate are not furnished by Rosemont Analytical.

Field weld pipe to adapter plate

3.5(89)

O.D. Ref

Pipe 3.00 in. Sched 40 Pipe Sleeve (Not by Rosemount Analytical) length by Customer

Masonry stack wall

35830005

2-7

Page 45
Image 45
Emerson Process Management IM-106-340 instruction manual Oxymitter 4000 Adapter Plate Installation