Oxymitter 4000
Instruction Manual
Electronic Assembly Replacement (Figure 9-5)
1.Remove the right housing cover uncovering the electronic assembly.
2.Depress and remove the J1 (cell and T/C) connector from the J1 socket. Loosen the three captive mounting screws (16, Figure
3.The J8 connector (heater leads) can be accessed by moving the J1 con- nector leads out of the slot on microprocessor board (17) and sliding the electronic assembly (12) partially out of the housing (Figure
4.Squeeze the J8 connector on the sides and carefully remove. The elec- tronic assembly can now be completely removed from the housing.
5.Reconnect the J8 connector to the power supply board. Make sure the connector is secure.
6.Holding the J1 connector leads, slide the electronic assembly the rest of the way into the housing. Align the electronic assembly so that it fits flush on the pins. To ensure that it is flush, gently try to rotate the elec- tronics. If the electronics rotates, repeat the alignment.
7.Reconnect the J1 connector to the microprocessor board. Ensure the connector is secure and tighten the three captive mounting screws on the microprocessor board (top board).
8.Replace the housing cover and ensure it is tight.
Terminal Block Replacement
1.Unscrew left housing cover (27, Figure
2.Loosen mounting screws (26) on terminal block (25) and carefully lift the block out of the housing.
3.Carefully align the new terminal block on the pins so that it sits flat in the housing. The round end of the terminal block should be on the opposite side of the housing conduit ports and should not be able to rotate.
4.Tighten the three mounting screws and ensure the terminal block is secure in the housing.
Fuse Replacement (Figure 9-7)
1.Remove the right housing cover uncovering the electronic assembly.
2.Depress and remove the J1 (cell and T/C) connector from the J1 socket. Loosen the three captive mounting screws (16, Figure
3.The J8 connector (heater leads) (Figure
4.Squeeze the J8 connector on the sides and carefully remove. The elec- tronic assembly can now be completely removed from the housing.
5.Completely remove the three mounting screws (16) on the microproces- sor board (17).