Oxymitter 4000

Temperatures

O2 Temp - Indicates the thermocouple temperature at the sensing cell; this should always be 1357°F (736°C).

O2 Temp Max - Maximum temperature the cell has seen. (Some process temperatures can exceed the 1357°F (736°C) setpoint temperature, and this will indicate this condition.)

Board Temp - The temperature inside the Oxymitter electronics housing 185°F (85°C), is the max.

Board Temp Max - This is the maximum temperature that the electronics has experienced over time.

 

Voltages

 

The raw mV signals feeding the temperature indications listed in the

 

previous paragraph.

 

Output Values

 

Indication of the current readings for O2 and mA.

LOI INSTALLATION

The LOI module connects to the top of the electronic assembly in the

 

electronics housing. There are four matching connectors (Figure 6-5)on the

 

back of the LOI module that allow the user to orient (rotate) the LOI as

 

desired.