Emerson Process Management IM-106-340 Startup and Operation of Oxymitter 4000 with LOI, Loi

Models: IM-106-340

1 176
Download 176 pages 38.09 Kb
Page 73
Image 73

Instruction Manual

IM-106-340, Rev. 4.0 May 2006

Oxymitter 4000

Section 6 Startup and Operation of Oxymitter 4000 with LOI

POWER UP

Power Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 6-1 Start Up Oxymitter 4000 Calibration . . . . . . . . . . . . . . . . . page 6-3 Navigating the Local Operator Interface . . . . . . . . . . . . . page 6-3 LOI Key Designations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 6-4 LOI Menu Tree . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 6-4 Oxymitter 4000 Setup at the LOI . . . . . . . . . . . . . . . . . . . . page 6-6 LOI Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 6-9 Oxymitter 4000 Test Points . . . . . . . . . . . . . . . . . . . . . . . . page 6-10 Remote Powered Loop LCD Display (Optional) . . . . . . . . page 6-10

Startup Display

When power is applied to the probe, the cell heater turns on. It takes approximately one half hour for the cell to heat to operating temperature. This condition is indicated by a "warm up" display on the LOI (Figure 6-1). This message will continue to display until the cell is up to operating temperature.

Operating Display

The normal operating display is the % O2 concentration. The "normal" display is shown in Figure 6-2.

Error

If there is an error condition at startup, an alarm message will be displayed. Refer to Section 8: Troubleshooting, to determine the cause of the error. Clear the error, cycle power, and the % O2 display should return.

LOI

The Local Operator Interface can be used to change the software and alarm settings, to adjust the high and low gas settings, and to initiate the calibration sequence. Refer to the LOI menu (Figure 6-4).

http://www.processanalytic.com

Page 73
Image 73
Emerson Process Management IM-106-340 instruction manual Startup and Operation of Oxymitter 4000 with LOI, Loi