Figure 9-9. Probe to Probe Head Assembly - Remote Electronics Only

O-RingProbe

Adapter

Housing

Board

 

Probe Signal

Wire Connector

4 Wire

2 Wire

Probe Heater

Wire Connector

37260058

Cell Replacement

This paragraph covers oxygen sensing cell replacement. Do not attempt to replace the cell until all other possibilities for poor performance have been considered. If cell replacement is needed, order the cell replacement kit (Table 10-1). Refer to Figure 9-3or Figure 9-4to view the component parts of the Oxymitter 4000.

The cell replacement kit (Figure 9-10)contains a cell and flange assembly, corrugated seal, setscrews, socket head cap screws, and anti-seize compound. The items are carefully packaged to preserve precise surface finishes. Do not remove items from the packaging until they are ready to be used. Spanner wrenches and hex wrenches needed for this procedure are part of an available special tools kit (Table 10-1and Figure 10-2).

9-16

Page 132
Image 132
Emerson Process Management IM-106-340 Probe to Probe Head Assembly Remote Electronics Only, Cell Replacement