Figure 8-13. Fault 11, Bad Cell

 

HEATER T/C

 

DIAGNOSTIC

HEATER

SW2

ALARMS

O2 CELL

ON

 

 

CALIBRATION

 

CALIBRATION RECOMMENDED

Oxymitter 4000

Fault 11, Bad Cell

Figure 8-13shows the electronic assembly for an Oxymitter 4000 with a membrane keypad (upper view) and an Oxymitter 4000 with an LOI (lower view).

Membrane Keypad

When Fault 11 is detected, the O2 CELL flashes three times, pauses for three seconds, and repeats.

TEST

POINTS

INC INC

HIGH LOW GAS GAS

DEC DEC

O2 CELL mV +

O2 CELL mv - HEATER T/C + HEATER T/C -

CAL

TEST GAS + PROCESS - % O2

TP1 TP2 TP3 TP4

TP5

TP6

J1 RED YEL GRN ORG

1.The bad cell alarm activates when the cell exceeds the maximum resistance value.

2.The cell should be replaced. See "Cell Replacement" in Section 9: Maintenance and Service, for cell replacement instructions.

KEYPAD

Alarms

O2 Cell Bad

LOI

37260029

 

LOI

When Fault 11 is detected, the LOI displays the "O2 Cell Bad" message. Refer to the comments and procedures in paragraphs 1 and 2 above.