Oxymitter 4000

Instruction Manual

IM-106-340, Rev. 4.0 May 2006

Table 8-1. Diagnostic/Unit

Alarm Fault Definitions -

Membrane Keypad Only

LED

Flashes

Status

4-20 mA Line

Fault

Recoverable

HEATER T/C

1

Open

3.5 mA (factory default)*

1

No

 

2

Shorted

3.5 mA (factory default)*

2

No

 

3

Reversed

3.5 mA (factory default)*

3

No

 

4

A/D Comm Error

3.5 mA (factory default)*

4

No

 

 

 

 

 

 

HEATER

1

Open

3.5 mA (factory default)*

5

No

 

2

High High Temp

3.5 mA (factory default)*

6

No

 

3

High Case Temp

3.5 mA (factory default)*

7

Yes

 

4

Low Temp

3.5 mA (factory default)*

8

Yes

 

5

High Temp

3.5 mA (factory default)*

9

Yes

O2 CELL

1

High mV

3.5 mA (factory default)*

10

Yes

 

3

Bad

Track O2

11

Yes

 

4

EEprom Corrupt

3.5 mA (factory default)*

12

No

CALIBRATION

1

Invalid Slope

Track O2

13

Yes

 

2

Invalid Constant

Track O2

14

Yes

 

3

Last Calibration Failed

Track O2

15

Yes

 

**

Calibration Recommended

Track O2

 

Yes

*Critical alarm conditions will render the O2 measurement as unusable, and any of these events will cause the 4-20 mA signal to go to a user-selectable limit of 3.5 mA or 21.6 mA (position 3 of SW2). Factory default value is 3.5 mA. Alarms which are not self-clearing (Self-Clearing = NO) will require a reset. Perform the Reset Procedure in Section 3: Configuration of Oxymitter 4000 with Membrane Keypad to continue operation.

**The CALIBRATION RECOMMENDED alarm flashes the Calibration Recommended LED on the operator's keypad.

Table 8-2. Diagnostic/Unit

Alarm Fault Definitions - LOI

Message

Status

Fault Number

Self Clearing

O2 T/C OPEN

Heater T/C Open

1

No

O2 T/C SHORTED

Heater T/CShorted

2

No

O2 T/C REVERSED

Heater T/C Polarity Reversed

3

No

 

 

 

 

ADC ERROR

A/D Comm Error

4

No

O2 HEATER OPEN

O2 Heater Open

5

No

VERY HI O2 TEMP

Very High Process Temperature

6

No

BOARD TEMP HI

Electronics Overheated

7

Yes

O2 TEMP LOW

Low Process Temperature

8

Yes

O2 TEMP HI

High Process Temperature

9

Yes

 

 

 

 

O2 CELL OPEN

O2 Cell Open

10

Yes

O2 CELL BAD

O2 Cell Failed

11, 13, 14

Yes

EEPROM CORRUPT

EEprom Failed

12

No

CALIB FAILED

Last Calibration Failed

15

Yes

LINE FREQ ERROR

Incorrect Input Line Frequency Detected on Power Up

 

No

8-6