5 - 10

FUNCTION REFERENCE

AUTOSET

Description:

The AUTOSET function sets the oscilloscope so that an optimum display of the input signals is obtained within the same mode analog or digital. Operating the AUTOSET key results in:

-Channels with an input signal are switched on; others are switched off.

-Input coupling is set to ac; because of this autoset does not function at very low signal frequencies.

-Input impedance is set to 1 MΩ

-Input attenuator settings are optimized; VAR function off.

-Bandwidth limiter and INVert are switched off.

-ALT or CHOP mode selected to most optimal display.

-Vertical POS selection in center screen.

-Edge triggering on positive slope is activated.

-Trigger source is the channel with the lowest frequency; at equal frequency the lower channel number is selected. Trigger source is EXT TRIG input when a signal is present, and Trigger View becomes active.

-Trigger coupling is set to ac and level-pp on.

-Horizontal mode is switched to MAIN TB only.

-Cursors and usertext are switched off; settings display is switched on.

These settings are suitable for most signal conditions. You can also customize AUTOSET to your own application or preference. For information about AUTOSET programming, refer to AUTOSET USERPROG.

Key sequence:

AUTOSET

Key to start the autoset

Remote commands:

CPL: AS (Command for Auto Setup)

Refer to Chapter 6 for full details.