6 - 50

THE CPL PROTOCOL

6.8 SETUP

The Query Setup (QS) and Program Setup (PS) commands can be used together in order to retrieve and restore a complete or partical setup of the oscilloscope.

When a setup is requested from the oscilloscope, it will send its setup as a sequence of strings. Each separate string describes a setup node. By adding a parameter to the QS command a particular setup node can be queried.

So the query program could be:

PRINT #1,"QS"

Complete setup

PRINT

#1,"QS",2

Only

node

2 (channel 2 settings)

PRINT

#1,"QS",S

Only

node

S, where S must be one of the

 

 

values specified below.

Node numbers (S) can have the following values:

hex

dec

meaning

01

1

Channel 1 settings

02

2

Channel 2 settings

03

3

Not used

04

4

Ext trig settings

 

 

(nodes above: attenuator, display on/off, coupling DC/AC, GND,

 

 

Invert, 50Ω/1 MΩ, continuous(var)/discrete, position control)

0E

14

Probe Scale settings (scale value, scale unit)

0F

15

Common vertical settings

 

 

(add 1+2, alt/chop, bandwidth limiter on/off, number of traces,

 

 

average, envelope).

10

16

Common horizontal settings

 

 

(X-deflection on/off, mode auto/trig/single/multiple, X-deflection

 

 

source, magnify, acquisition length, X-position.

11

17

Main Timebase settings

 

 

(timebase speed, triggering edge/TV, MTBI on/off, MAIN TB on/off,

 

 

trigger slope pos/neg, TV trigger F1/F2/line, noise suppression

 

 

on/off, continuous(var)/discrete, level-pp on/off, trigger source,

 

 

trigger coupling, TV trigger normal/hdtv)

12

18

Delayed Timebase settings

(timebase speed, trigger mode edge/tv, DEL’D TB on/off, edge trigger slope pos/neg, edge trigger source, trigger coupling)