Fluke PM3370B user manual Program Communication, 16THE CPL Protocol, Xonxoff, Hwl

Models: PM3370B

1 297
Download 297 pages 53.38 Kb
Page 233
Image 233

6 - 16THE CPL PROTOCOL

PROGRAM COMMUNICATION

PC

Purpose:

To program baudrate, parity mode, number of data and stopbits and the handshake method for computer communication.

After the command is sent, an <acknowledge> will be returned with the old communication parameters still active.

If the <acknowledge> = 0, the new communication parameters will be valid approximately 0.5 seconds later. The communication parameters are stored in battery backup memory and restored on power-up.

This command performs the same function as the UTILITY + RS232 SETUP menu.

Command:

"PC"

pds

baud

pds

parity

pds

data bits

 

 

 

to bypass handshake

 

 

pds

stop bits

pds

handshake

 

CR

<baud>

= 75, 110, 150, 300, 600, 1200, 2400, 4800, 9600, 19200,

 

 

38400

 

<parity>

= O, E or N

 

<data bits>

=

7 or 8

 

<stopbits>

=

1

 

<handshake>

=

XONXOFF

to enable Xon/Xoff handshake

 

 

HWL

to enable hardwareline handshake

 

 

If left out, both handshake methods are disabled.

Page 233
Image 233
Fluke PM3370B user manual Program Communication, 16THE CPL Protocol, Xonxoff, Hwl