HOW TO USE THE INSTRUMENT

4 - 67

DELAYED TIMEBASE

TRIGGER LEVEL

Just as is the case for Main Timebase triggering, proper triggering of the Delayed Timebase depends on the selection of the proper trigger level.

The Δ symbol in the Delayed Timebase trigger menu indicates the Δ control to be used to adjust the trigger level.

If the coupling for the vertical channel and the triggering are the same, a trigger level indicator is visible (D-).

If you have set up for triggered operation of the Del’dTB as described in the previous section, and the Del’dTB trace is not displayed, the trigger level needs adjustment.

Adjust the Δ control until the Del’dTB signal is visible.

1-

Δ

1- D-

D-

D: DTB TRIGGER LEVEL INDICATOR

ST6729

 

DELAYED TIMEBASE

COUPLING

The Del’dTB permits the same trigger coupling selection as for the MainTB (ac, dc, LF-rej, HF-rej). The level indication ’D-’ (for Del’dTB) is present only when the combination of the Del’dTB trigger coupling and the input coupling is useful.

Page 112
Image 112
Fluke PM3370B user manual Delayed Timebase Trigger Level, Delayed Timebase Coupling