GETTING STARTED

 

3 - 7

SIGNAL CRITERIA

ANALOG MODE

DIGITAL MODE

 

 

 

You need to make adjustments

Fastest

Slower

to the circuitry and watch

display

display

the signal change

update

update

Automatic measurements

Can’t use

Fully implemented

Signal

Math

Add, Subtract

All functions

 

Add, Subtract, Multiply

 

 

Signal

Analysis

Not available

Full analysis

 

Integration,

 

(optional)

 

Differentiation, FFT

 

 

Automatic Pass/Fail test

Not available

Fully implemented

 

 

 

(optional)

Autorange attenuator

Not available

Results in a displayed

 

 

 

signal with an ampli-

 

 

 

tude of 2 to 6.4 divi-

 

 

 

sions

Autorange timebase

Not available

Results in a signal

 

 

 

display of 2 to 6

 

 

 

waveform periods

ANALOG

1 2 3 4

RUN/STOP

ST6680

9312

Figure 3.4 Analog-Digital switching setup

Step 1 Press AUTOSET. The scope performs an AUTOSET in analog mode.

Step 2 Press the ANALOG key to change over to the digital mode. Check that the picture is identical to the one in the analog mode. The text ’DIGITAL MODE’ is displayed briefly at the bottom of the screen.