5 - 36

FUNCTION REFERENCE

Key sequence:

MATH MATH1(2)

TRACK

Control to select the filter function from the MATH 1 or MATH 2 menu.

Δ

Control to select the source signal.

on off

PARAM

DISPLAY SOURCE yes no

Toggle key to switch selected function on or off.

TRACK

Control to select the number of samples for the filter window.

Toggle key to switch the display of the source trace on or off.

ST6757

9303

Remote commands:

CPL: QW (Command to query a waveform)

Refer to Chapter 6 for full details.

GLITCH TRIGGER

Glitch triggering examines a single channel, on one or more time conditions.

Any one of the channel inputs can be used as trigger source and is selected with the keys ‘TRIG1, TRIG2’ or EXT TRIG. The same keys are used to toggle between positive and negative glitch triggering.

Time qualification is selected with softkeys. These time conditions are:

n>t1

triggers when the glitch is longer than the selected time t1. The

 

TRACK control is used to adjust t1.

n<t2

triggers when the glitch is shorter than the time t2. The TRACK

 

control is used to adjust t2.

range

triggers when the glitch duration is between two time limits as

 

specified by t1 and t2. The TRACK control is used to adjust t1 and

 

the control is used to adjust t2.

Page 156
Image 156
Fluke PM3370B user manual Glitch Trigger, Remote commands