HOW TO USE THE INSTRUMENT

4 - 43

TRACE STORAGE

SAVING MULTIPLE TRACES

Each memory location can store two channels.

-Connect the probe to channel 1.

-Press AUTOSET.

-Switch on channel 2.

-Turn the channel 2 POS control counterclockwise.

The upper trace displays the signal of channel 1, and the lower line represents channel 2.

-Press the SAVE key.

-Turn the TRACK control to select memory location ’m1’.

-Press the ’save’ softkey to store two traces in ’m1’.

Observe that m1 was already filled, so the confirmation submenu is displayed.

-Press the softkey next to ’yes’.

This overwrites memory ’m1’ with the new traces.

TRACE STORAGE

RECALLING MULTIPLE TRACES

-Press the RECALL menu key.

The memory location is already preset to the last saved (m1).

-Press the ’DISPLAY on off’ softkey so that ’m1’ is displayed.

-Turn the Δ control counterclockwise.

Observe that both traces are displayed and shifted to the lower screen.

-Press the softkey next to ’trace register’ to select ’trace’.

Observe that the menu displays both traces of register 1 (m1.1 and m1.2). Both are filled: m1.1 with the signal of channel 1 and m1.2 with the signal from channel 2.

-Turn the TRACK control to select m1.1.

-Press the ’DISPLAY on off’ softkey to off.

Observe that the stored trace of channel 1 is not displayed any more.

-Press the ’CLEAR DISPLAY’ softkey to clear all stored traces before continuing to the next section.

Page 88
Image 88
Fluke PM3370B Trace Storage Saving Multiple Traces, Trace Storage Recalling Multiple Traces, Press the Recall menu key