THE CPL PROTOCOL

6 - 15

LOCAL LOCKOUT

LL

Purpose:

This instruction will inhibit the Go to Local function of the STATUS/LOCAL key on the frontpanel.

Once activated, the Local Lockout State is disabled by sending the Go to Local (GL), the Reset Instrument (RI) command or by cycling power OFF and ON. (Refer also to "Go to Remote" and "Go to Local").

There is no front panel equivalent for this command.

Command:

"LL"

CR

Response:

acknowledge

CR

Example:

100 PRINT #1,"LL" :’Send Command

110 GOSUB 1000 :’Sync on acknowledge

Local, Remote, Remote+Local Lockout States

power-on

Local State

"Go to Remote"

"Go to Local" "Reset Instrument"

<STATUS/LOCAL button>

Remote State

"Local Lockout"

"Go to Local" "Reset Instrument"

Remote with

Local Lockout

State

Page 232
Image 232
Fluke PM3370B user manual Local Lockout