FUNCTION REFERENCE

5 - 75

TIME MEASUREMENTS

Description:

Time measurements can be made using the cursors or using automated, calculated measurement routines. Cursors operate in the analog mode as well as in the digital mode.

Calculated time measurements are available in the digital mode only. Using the cursors, three time interval readouts can be selected:

-ΔT gives the time interval (seconds) between the cursors.

-1ΔT gives an inverted readout of the time interval, and is read in Hz.

-ΔT-ratio gives a percentage (%) that can be reset by using the ΔT=100% softkey. For more details, refer to CURSORS READOUT.

In the digital mode, the MEASURE menu accesses automatic time measurements. This menu is reached via the key sequence ‘MEASURE >> MEAS 1(2) >> time’. The TRACK control us used to select the type of measurement and the Δ control is used to select the waveform.

The result of the measurement is displayed in the top left corner of the screen.

 

PERIOD

 

PULSE

 

DISTAL

 

MESIAL

1

PROXIMAL

 

RISE

ST6744

FALL

The following measurements are possible:

(Refer to section MEASURE MENU for details of signal parameters).

-freq (FREQUENCY).

Measures the reciprocal of the PERIOD of the signal. Frequency is calculated between the first and third mesial of the signal. At least one complete signal period must be acquired to get a valid measurement.

freq =

1

time of 3rd mesial - time 1st mesial

Page 195