THE CPL PROTOCOL

6 - 19

PROGRAM TEXT

PT

Purpose:

To program text to an oscilloscope.

If the S parameter is specified, setup text is programmed. The text will be set into one of the setup registers of the oscilloscope. The parameter n specifies the setup register. A maximum number of 22 characters is allowed. The remainder of the text field is set to ’spaces’.

If the S parameter is not specified, user text is programmed. The text will be displayed on the screen of the oscilloscope. A maximum number of 64 characters is allowed. The remainder of the text field is set to ’spaces’. The parametern may not be used.

The programmed text can be read with the Query Text (QT) command.

Command 1:

"PT"

pds

"S"

pds

n

CR

Response 1:

acknowledge

CR

Command 2:

char

CR

Response 2:

acknowledge

CR

Page 236
Image 236
Fluke PM3370B user manual Program Text, Pds, Char