5 - 82

FUNCTION REFERENCE

Source and slope are selected with the same TRIG 1, TRIG .., and EXT TRIG keys that are used for the main timebase trigger source and slope selection. The delayed timebase source and slope have their own readout. This readout can be found at the bottom right corner of the screen, below the readout for the main timebase trigger source and slope.

If the channel and trigger coupling are the same (e.g., both ac or both dc), the trigger level is indicated on the screen by D-.

For more details refer to the DEL’D TB section.

Key sequence:

 

DEL'D TB

DTB

on off

 

Toggle softkey to switch DEL’D TB ’on’.

starts

trig'd

Toggle softkey to select the DEL’D TB trigger mode.

Δ

Control to adjust the DEL’D TB trigger level.

ac dc lf-rej

hf-rejToggle softkey to select the DEL’D TB trigger coupling modes ac, dc, lf- reject and hf- reject.

ST6774

9303

TRIGGER LEVEL

Description:

LEVEL selects the signal level at which the timebase will trigger. For triggering, the level must be within the peaks of the signal.

TRIGGER LEVEL sets the trigger level for the main timebase. In level-pp ’on’ mode, the level range is clamped within the peak-peak values of the signal. Then the MTB will always be triggered as shown in the following figure. When level-pp is ’off’, the level range is from -8 ... +8 divisions.

Leveling in DEL’D TB is adjusted with the Δ control. It is activated in the triggered DEL’D TB mode. The range is from -8 ... +8 divisions.

Page 202
Image 202
Fluke PM3370B user manual Toggle softkey to switch DEL’D TB ’on’