THE CPL PROTOCOL

6 - 23

wave_nr The oscilloscope waveform destination:

011 - 084 for m1.1 - m8.4 01e - 08e for m1.e - m8.e 091 - 504 for m9.1 - m50.4 (Extended Memory only)

09e - 50e for m9.e - m50.e (Extended Memory only)

(xx.e for 2 ch. models only; xx.3 and xx.4 for 4 ch. models only)

admin 16 parameters, separated by a "pds":

 

 

 

 

ACTION BY

 

PARAMETER

TYPE

EXAMPLE

OSCILLOSCOPE NOTES

 

 

 

 

 

 

trace_name

 

string

m4.1

ignored

 

Y_unit

 

string

V

interpreted

 

X_unit

 

string

s

interpreted

 

Y_zero

 

number

+3

interpreted

 

X_zero

 

number

-8.625E-6

ignored

 

Y_resolution

number

78.13E-3

interpreted

 

X_resolution

number

1E-6

interpreted

 

Y_range

 

number

65535

ignored

 

date

 

string

25-01-1994

interpreted

3)

time

 

string

17:10:25:40

interpreted

4)

dT-corr

 

number

375E-3

interpreted

1)

min/max

 

number

0

interpreted

2)

mult_shot_tot

number

32

interpreted

 

mult_shot_nr

number

5

interpreted

 

reserved

 

string

0

ignored

 

reserved

 

string

0

ignored

 

Notes:

1)

dT = dT-corr * X-resolution

0 = waveform trace

 

2)

1 = min/max trace envelopes

 

3)

date format:

European = E25-01-1994

 

 

 

 

American = A01-25-1994

 

 

 

 

Japanees = J1994-01-25

 

for compatibility reasons (initial versions) European format without leading character can be used.

4) for compatibility reasons (initial versions) hundreds of seconds can be omited

count The number of samples in the waveform: 512, 1024, 2048, 4096, 8192, 16384 or 32768 (Extended Memory: 2048 and higher)

Page 240
Image 240
Fluke PM3370B user manual Action by Parameter Type Example Oscilloscope Notes