FUNCTION REFERENCE

5 - 41

Key sequence:

ON

Toggle key to switch a channel on/off.

TRIG VIEW

Toggle key to switch the display of the External Trigger signal on/off.

AC/DC/GND

Toggle key for vertical input coupling of the input channels.

AC/DC

Toggle key for vertical input coupling of the External

Trigger input.

INPUT IMPEDANCE (200 MHZ MODELS ONLY)

Description:

For high-frequency measurements the input impedance can be switched from 1 MΩ to 50Ω. Input impedance is automatically adapted to the type of probe. Manual switching is done with the toggle key 50Ω in the VERT MENU menu. The 50Ω input is not available for the EXT TRIG.

The input is protected in the 1 MΩ as well as the 50Ω setting:

-1 MΩ: protected for signals up to 400V.

-50Ω: protected for peak voltages up to 50V or up to 5V rms.

For details, refer to the Reference Manual.

Key sequence:

Toggle key to switch 50Ω input impedance of CH1.

..

Toggle key to switch 50Ω input impedance of CH..